High-sensitivity cardiac troponin t concentrations below the limit of detection to exclude acute myocardial infarction: a prospective evaluation

Initial reports suggest that concentrations of high-sensitivity cardiac troponin T (hs-cTnT) (Roche Diagnostics Elecsys(®)) below the limit of blank (LoB) (3 ng/L) or limit of detection (LoD) (5 ng/L) of the assay have almost 100% negative predictive value (NPV) for acute myocardial infarction (AMI)...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical chemistry (Baltimore, Md.) Md.), 2015-07, Vol.61 (7), p.983-989
Hauptverfasser: Body, Richard, Burrows, Gillian, Carley, Simon, Cullen, Louise, Than, Martin, Jaffe, Allan S, Lewis, Philip S
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Initial reports suggest that concentrations of high-sensitivity cardiac troponin T (hs-cTnT) (Roche Diagnostics Elecsys(®)) below the limit of blank (LoB) (3 ng/L) or limit of detection (LoD) (5 ng/L) of the assay have almost 100% negative predictive value (NPV) for acute myocardial infarction (AMI), particularly among patients without electrocardiograph (ECG) evidence of ischemia. We aimed to prospectively validate those findings. We included adults presenting to the emergency department with suspected cardiac chest pain. Standard troponin T (cTnT) and hs-cTnT (both Roche Elecsys) were tested in samples drawn on arrival. The primary outcome was AMI, adjudicated by 2 investigators on the basis of clinical data and ≥12-h cTnT testing. We also evaluated diagnostic performance when AMI was readjudicated on the basis of hs-cTnT (≥12-h) concentrations. Of 463 patients included, 79 (17.1%) had AMI. Twenty-four patients (5.2%) had hs-cTnT concentrations below the LoB, although none had AMI. Ninety-six patients (20.7%) had hs-cTnT concentrations below the LoD, 1 of whom had AMI. Thus, diagnostic sensitivity was 98.7% (95% CI 87.5%-98.6%) and NPV was 99.0% (95% CI 94.3%-100.0%). Of the 17.3% (n = 80) patients with hs-cTnT below the LoD and no ECG ischemia, none had AMI. Thus, diagnostic sensitivity was 100.0% (95% CI 95.4%-100.0%) and NPV was 100.0% (95% CI 95.5%-100.0%). Sensitivity and NPV were maintained when AMI was readjudicated on the basis of hs-cTnT. Our findings confirm that patients with nonischemic ECG and undetectable hs-cTnT at presentation have a very low probability of AMI, although the proportion of patients affected was smaller than in previous research.
ISSN:0009-9147
1530-8561
DOI:10.1373/clinchem.2014.231530