Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

In external beam radiotherapy treatment planning for patients with thoracic malignancies, respiratory‐correlated CT (4D CT) is used to obtain high quality studies in the presence of respiratory motion. When helical 4D CT scans are acquired with a Brilliance CT Big Bore, the pitch must meet two condi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of applied clinical medical physics 2015-05, Vol.16 (3), p.389-398
Hauptverfasser: Hilgers, Guido, Nuver, Tonnis, Minken, André
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In external beam radiotherapy treatment planning for patients with thoracic malignancies, respiratory‐correlated CT (4D CT) is used to obtain high quality studies in the presence of respiratory motion. When helical 4D CT scans are acquired with a Brilliance CT Big Bore, the pitch must meet two conditions. It must be low enough to avoid motion artifacts, and high enough to cover the entire scan length within 120 s to prevent overheating of the X‐ray tube. We developed a nomogram that can be used to obtain a suitable pitch satisfying both requirements. We also assessed the effects on the image quality of a pitch that exceeds the maximum pitch, and of a field of view (FOV) reduction. It was shown that, for AV G and MIP reconstructions, the manufacturer's maximum pitch equation yields an underestimation due to its FOV term. PACS number: 87.57.Q‐, 87.57.cp