Swallowing accelerometry signal feature variations with sensor displacement
Highlights • First study to systematically evaluate the effect of positional offsets of sensor location on dualaxes swallowing accelerometry signals. • Sensor can be displaced 4 mm inferior to and 6 mm superior to the baseline position (midline on neck, below thyroid cartilage) without significant...
Gespeichert in:
Veröffentlicht in: | Medical engineering & physics 2015-07, Vol.37 (7), p.665-673 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Highlights • First study to systematically evaluate the effect of positional offsets of sensor location on dualaxes swallowing accelerometry signals. • Sensor can be displaced 4 mm inferior to and 6 mm superior to the baseline position (midline on neck, below thyroid cartilage) without significantly altering time-frequency signal features. • Sensor can be displaced 4 mm lateral to the baseline position without significantly altering timefrequency signal features. • Time-frequency features are more robust to positional offsets of sensor than time domain and frequency domain features. • Classification of swallowing conditions based on swallow accelerometry should consider time-frequency features if positional robustness of sensor placement is important. |
---|---|
ISSN: | 1350-4533 1873-4030 |
DOI: | 10.1016/j.medengphy.2015.04.007 |