Measuring Tissue Back-Pressure - In Vivo Injection Forces During Subcutaneous Injection
Purpose Limited information is available on injection forces of parenterals representing the in vivo situation. Scope of the present study was to investigate the contribution of the subcutaneous (sc) tissue layer to injection forces during in vivo injection. Methods Göttingen minipigs received injec...
Gespeichert in:
Veröffentlicht in: | Pharmaceutical research 2015-07, Vol.32 (7), p.2229-2240 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
Limited information is available on injection forces of parenterals representing the
in vivo
situation. Scope of the present study was to investigate the contribution of the subcutaneous (sc) tissue layer to injection forces during
in vivo
injection.
Methods
Göttingen minipigs received injections of isotonic dextran solutions (1–100 mPas) into the
plica inguinalis
using different injection rates and volumes (0.025–0.2 mL/s and 2.5 vs. 4.5 mL).
Results
The contribution of the sc back-pressure to injection forces was found to increase linearly with viscosity and injection rate ranging from 0.6 ± 0.5 N to 1.0 ± 0.4 N (1 mPas), 0.7 ± 0.2 N to 2.4 ± 1.9 N (10 mPas), and 1.8 ± 0.6 N to 4.7 ± 3.3 N (20 mPas) for injection rates of 0.025 to 0.2 mL/s, respectively. Variability increased with viscosity and injection rate. Values are average values from 10 randomized injections. A maximum of 12.9 N was reached for 20 mPas at 0.2 mL/s; 6.9 ± 0.3 N was determined for 100 mPas at 0.025 mL/s. No difference was found between injection volumes of 2.5 and 4.5 mL. The contribution of the tissue was differentiated from the contribution of the injection device and a local temperature effect. This effect was leading to warming of the (equilibrated) sample in the needle, therefore smaller injection forces than expected compensating tissue resistance to some parts.
Conclusions
When estimating injection forces representative for the
in vivo
situation, the contribution of the tissue has to be considered as well as local warming of the sample in the needle during injection. |
---|---|