Revision of Recalled Modular Neck Rejuvenate and ABG Femoral Implants

Abstract Modular neck femoral stems have been associated with adverse local tissue reactions (ALTR), leading to a voluntary recall, but these effects have not been well-characterized. A retrospective review of intraoperative findings and cobalt/chromium levels was performed in 103 hips undergoing re...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2015-05, Vol.30 (5), p.822-826
Hauptverfasser: Walsh, Christopher P., MD, Hubbard, James C., BS, Nessler, Joseph P., MD, Markel, David C., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Modular neck femoral stems have been associated with adverse local tissue reactions (ALTR), leading to a voluntary recall, but these effects have not been well-characterized. A retrospective review of intraoperative findings and cobalt/chromium levels was performed in 103 hips undergoing revision for ALTR. The average preoperative serum cobalt level was 7.6 μg/L (range 1.1–23 μg/L) and chromium level was 1.8 μg/L (range 0.1–6.8 μg/L). Metallic sludge was noted in 100%, synovitis in 98%, pericapsular rind in 82%, and calcar erosion in 85%. An osteotomy was required for removal in 44%. We concluded that revision of modular neck femoral stems is associated with increased preoperative metal ion levels and stem-neck corrosion. Despite advanced stem explantation techniques, osteotomy was frequently required, leading to increased morbidity.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2014.12.002