Changing the paradigm of laboratory quality control through implementation of real-time test results monitoring: For patients by patients
To develop and implement a quality control protocol using real-time patient data with immediate failure analysis and prevention of releasing results that exceed the allowable total error. Patient data are analyzed in real time using algorithms that incorporate moving medians and moving means for sel...
Gespeichert in:
Veröffentlicht in: | Clinical biochemistry 2015-05, Vol.48 (7-8), p.508-513 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To develop and implement a quality control protocol using real-time patient data with immediate failure analysis and prevention of releasing results that exceed the allowable total error.
Patient data are analyzed in real time using algorithms that incorporate moving medians and moving means for selected chemistry analytes. Simulation software was developed to determine optimal algorithms, establish error limits, and number of patient results for calculation of a single cumulative datum point. Algorithms for moving median (MovMed) and mean (MovMen) were chosen and validated for each analyte. Error limits (TEa) were established using biological and analytical variation with a goal of greater than 90% error detection rate during simulation runs. Middleware software was developed to prohibit the release of patient results upon error detection.
A block size of 50 was determined to be the optimal number of patient results used in cumulative calculations. The application of MovMed and MovMen algorithms achieved 0% false rejection for 24 out of 28 tests (85.7%) during the simulation phase. Four tests had a false rejection rate ranging from 0.2 to 1.0%. Error detection rates of 100% were achieved for 16 out of 28 tests (57.1%). Twelve tests had error detection rates ranging from 94.5 to 99.8%. Traditional QC material utilization was reduced by approximately 75–85% and repeat analysis was reduced by approximately 50%.
We successfully developed and implemented a real-time quality control protocol using patient results with true error detection and without release of erroneous results.
•We show how to develop and implement real-time quality control protocols based on patient test results.•The algorithms use moving means and moving medians of patient results.•These protocols allow continuous instrument monitoring and permit timely operator intervention.•This avoids non-commutability of conventional liquid quality control materials, achieve significant cost savings. |
---|---|
ISSN: | 0009-9120 1873-2933 |
DOI: | 10.1016/j.clinbiochem.2014.12.016 |