Electrochemical determination of the anticancer drug taxol at a ds-DNA modified pencil-graphite electrode and its application as a label-free electrochemical biosensor

In this study a novel biosensor for determination of taxol is described. The interaction of taxol with salmon-sperm double-stranded DNA (ds-DNA) based on the decreasing of the oxidation signals of guanine and adenine bases was studied electrochemically with a pencil-graphite electrode (PGE) using a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Talanta (Oxford) 2015-03, Vol.134, p.60-64
Hauptverfasser: Tajik, Somayeh, Taher, Mohammad Ali, Beitollahi, Hadi, Torkzadeh-Mahani, Mosoud
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this study a novel biosensor for determination of taxol is described. The interaction of taxol with salmon-sperm double-stranded DNA (ds-DNA) based on the decreasing of the oxidation signals of guanine and adenine bases was studied electrochemically with a pencil-graphite electrode (PGE) using a differential pulse voltammetry (DPV) method. The decreases in the intensity of the guanine and adenine oxidation signals after interaction with taxol were used as indicator signals for the sensitive determination of taxol. DPV exhibits a linear dynamic range of 2.0×10−7–1.0×10−5M for taxol with a detection limit of 8.0×10−8M. Finally, this modified electrode was used for determination of taxol in some real samples. A taxol DNA biosensor using ds-DNA modified pencil graphite electrode based on guanine and adenine signals was constructed. [Display omitted] •A taxol DNA biosensor was constructed.•DPV exhibits a linear dynamic range from 2.0×10−7 to 1.0×10−5M.•This modified electrode was used for determination of taxol in real samples.
ISSN:0039-9140
1873-3573
DOI:10.1016/j.talanta.2014.10.063