Effect of microneedles shape on skin penetration and minimally invasive continuous glucose monitoring in vivo

•Effect of microneedle shape on skin penetration and minimally invasive continuous glucose monitoring investigated.•Straight silicon microneedle arrays (SSMA) and tapered silicon microneedle arrays (TSMA) were fabricated using standard silicon micromachining techniques.•Mode-I planar crack and mode-...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Sensors and actuators. A. Physical. 2013-12, Vol.203, p.373-381
Hauptverfasser: Chua, Beelee, Desai, Shashi P., Tierney, Michael J., Tamada, Janet A., Jina, Arvind N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:•Effect of microneedle shape on skin penetration and minimally invasive continuous glucose monitoring investigated.•Straight silicon microneedle arrays (SSMA) and tapered silicon microneedle arrays (TSMA) were fabricated using standard silicon micromachining techniques.•Mode-I planar crack and mode-II ring crack propagation were assumed for TSMA and SSMA penetration mechanisms, respectively.•Sensor currents from the applied CGM system prototypes ranged from 100nA to 400nA and they appeared to follow the trend of the reference blood glucose. The effect of microneedle shape on skin penetration and minimally invasive continuous glucose monitoring (CGM) was successfully investigated. Straight silicon microneedle arrays (SSMA) and tapered silicon microneedle arrays (TSMA) were fabricated using standard silicon micromachining techniques. Microneedle heights were ∼325μm and ∼350μm and pitch was 400μm and 450μm, respectively. Mode-I planar crack and mode-II ring crack propagation were assumed for TSMA and SSMA penetration mechanisms, respectively. Skin penetration was first qualitatively evaluated via methylene blue staining. TSMA and SSMA were also mounted on CGM system prototypes and applied on human subjects with an impact velocity of 10m/s and worn for 6.5h. The hollow microneedles allowed the glucose from the interstitial fluid in the epidermis to diffuse into the sensor chamber of the CGM system prototype. Sensor currents from the applied CGM system prototypes ranged from 100nA to 400nA and they appeared to follow the trend of the reference blood glucose sensor that ranged from ∼90 to 220mg/dL. The CGM system prototype with SSMA yielded lower sensor current than that with TSMA. This may be due to the SSMA preferential penetration of skin via mode-II ring crack propagation, resulting in possible lumen occlusion from displaced skin material.
ISSN:0924-4247
1873-3069
DOI:10.1016/j.sna.2013.09.026