Use of the Amplatzer Vascular Plug II Device to Occlude Different Types of Patent Ductus Arteriosus in Pediatric Patients
Introduction Transcatheter patent ductus arteriosus (PDA) closure is difficult in small patients in part due to the diverse PDA morphology. We describe a multicenter pediatric experience using the Amplatzer Vascular Plug II device (AVP II) to occlude PDA. Methods All patients undergoing transcathete...
Gespeichert in:
Veröffentlicht in: | Journal of interventional cardiology 2015-04, Vol.28 (2), p.198-204 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
Transcatheter patent ductus arteriosus (PDA) closure is difficult in small patients in part due to the diverse PDA morphology. We describe a multicenter pediatric experience using the Amplatzer Vascular Plug II device (AVP II) to occlude PDA.
Methods
All patients undergoing transcatheter closure of PDA with AVP II from April 2008 until May 2012 were included. Clinical, angiographic, and echocardiographic data were collected.
Results
One hundred and seventy‐seven procedures were performed. Median age was 14 months (2–180) with median weight 9.7 kg (4.7–68). The morphological PDA classification was Type A in 66 (37.1%), Type C in 27 (15.3%), Type D in 17 (9.6%), and Type E in 67 (37.9%). The median minimum PDA diameter was 2.6 mm (1.2–7.7 mm). The implanted device sizes were: 4 mm in 17 patients (9.6%), 6 mm in 81 (45.8%), 8 mm in 56 (31.6%), 10 mm in 18 (10.2%), and 12 mm in 5 (2.8%). The implanted device was mean of 2.6 + 0.7 times the ductus narrowest diameter and mean of 1 + 0.5 times the ductus largest diameter. Complications included: one severe left pulmonary artery stenosis and one device embolization. No aortic obstruction occurred. Closure was complete in the 175 remaining patients.
Conclusions
The AVP II is an effective and safe device for PDA closure. It is particularly useful when dealing with nontypical PDA shapes and in small infants where it eliminated the risk of device‐related aortic obstruction. The AVPII is an addition to the PDA device closure armamentarium in pediatric patients. |
---|---|