Benchmark IMRT evaluation of a Co-60 MRI-guided radiation therapy system

Abstract A device for MRI-guided radiation therapy (MR-IGRT) that uses cobalt-60 sources to deliver intensity modulated radiation therapy is now commercially available. We investigated the performance of the treatment planning and delivery system against the benchmark recommended by the American Ass...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiotherapy and oncology 2015-03, Vol.114 (3), p.402-405
Hauptverfasser: Wooten, H. Omar, Rodriguez, Vivian, Green, Olga, Kashani, Rojano, Santanam, Lakshmi, Tanderup, Kari, Mutic, Sasa, Li, H. Harold
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract A device for MRI-guided radiation therapy (MR-IGRT) that uses cobalt-60 sources to deliver intensity modulated radiation therapy is now commercially available. We investigated the performance of the treatment planning and delivery system against the benchmark recommended by the American Association of Physicists in Medicine (AAPM) Task Group 119 for IMRT commissioning and demonstrated that the device plans and delivers IMRT treatments within recommended confidence limits and with similar accuracy as linac IMRT.
ISSN:0167-8140
1879-0887
DOI:10.1016/j.radonc.2015.01.015