Late failure of annealed highly cross-linked polyethylene acetabular liner

Highly cross-linked polyethylene (HXPE) in total hip arthroplasty (THA) has been shown to significantly decrease wear rates compared with conventional polyethylene (CPE). However, crosslinking, thermal treatment and oxidation can decrease the mechanical properties of PE, and several cases of fractur...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the mechanical behavior of biomedical materials 2013-12, Vol.28, p.206-212
Hauptverfasser: Hara, Daisuke, Nakashima, Yasuharu, Yamamoto, Takuaki, Higashihara, Shinshichiro, Todo, Mitsugu, Hirata, Masanobu, Akiyama, Mio, Iwamoto, Yukihide
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highly cross-linked polyethylene (HXPE) in total hip arthroplasty (THA) has been shown to significantly decrease wear rates compared with conventional polyethylene (CPE). However, crosslinking, thermal treatment and oxidation can decrease the mechanical properties of PE, and several cases of fracture of remelted HXPE liners were reported. We present, for the first time, unexpected failures of THA with the use of annealed HXPE liners in two patients occurring at 7 and 8 years after operation. Operative findings revealed dislocated liners from the metal shell and a fracture of the superior rim at the rim–dome junction in both liners. Scanning electron microscopy showed that the cracks initiated at the rim and propagated toward the articular surface. Both liners showed generally a low amount of oxidation (less than 1.00) at the articular surface and low wear rates; however, oxidation at the rim was relatively higher (mean 1.55). These findings suggested that decreased mechanical properties at the rim–dome junction due to cross-linking, annealing and oxidation might have been caused breakage of the HXPE liners after a long implantation time, although the annealed HXPE achieved low degree of wear. [Display omitted] •We present for the first time two failures of annealed and cross-linked PE.•Both cases show broken and avulsed rim at the rim–dome junction.•SEM shows that the breakage initiated at the rim and propagated to inner surface.•Oxidation at the rim was relatively higher and might reduce mechanical properties.•Reduced mechanical properties can explain the breakage of annealed and remelted PE.
ISSN:1751-6161
1878-0180
DOI:10.1016/j.jmbbm.2013.08.003