Estimation and Optimization of the Principal H-Maser Parameters
The scale along the coordinate axes in the equations of motion hitherto used for an H-maser (in relative units) is shown to depend on Q as the quality factor for an H-maser. This hinders both analysis of the operation of the H-maser and optimization of its parameters. The modified equations of motio...
Gespeichert in:
Veröffentlicht in: | Measurement techniques 2004-06, Vol.47 (6), p.562-567 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The scale along the coordinate axes in the equations of motion hitherto used for an H-maser (in relative units) is shown to depend on Q as the quality factor for an H-maser. This hinders both analysis of the operation of the H-maser and optimization of its parameters. The modified equations of motion proposed here do not have that shortcoming. We determine the atomic flux into the bulb that gives the H-maser the best metrological characteristics.[PUBLICATION ABSTRACT] |
---|---|