Estimation and Optimization of the Principal H-Maser Parameters

The scale along the coordinate axes in the equations of motion hitherto used for an H-maser (in relative units) is shown to depend on Q as the quality factor for an H-maser. This hinders both analysis of the operation of the H-maser and optimization of its parameters. The modified equations of motio...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Measurement techniques 2004-06, Vol.47 (6), p.562-567
1. Verfasser: Rusin, F S
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The scale along the coordinate axes in the equations of motion hitherto used for an H-maser (in relative units) is shown to depend on Q as the quality factor for an H-maser. This hinders both analysis of the operation of the H-maser and optimization of its parameters. The modified equations of motion proposed here do not have that shortcoming. We determine the atomic flux into the bulb that gives the H-maser the best metrological characteristics.[PUBLICATION ABSTRACT]
ISSN:0543-1972
1573-8906
DOI:10.1023/B:METE.0000039760.95024.66