Determining the Instantaneous Effective Energy of x-Ray Tube Bremsstrahlung

A method is described for determining the instantaneous effective energy of x-ray tube brehmsstrahlung by means of two semiconductor detectors employing epitaxial GaAs structures and a measurement circuit, which together determine the effective energy with an error of 5% in the range 20-80 keV in th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Measurement techniques 2003-08, Vol.46 (8), p.806-809
Hauptverfasser: Dvoryankin, V F, Dikaev, Yu M, Kudryashov, A A, Sokolovskii, A A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method is described for determining the instantaneous effective energy of x-ray tube brehmsstrahlung by means of two semiconductor detectors employing epitaxial GaAs structures and a measurement circuit, which together determine the effective energy with an error of 5% in the range 20-80 keV in the presence of nonlinearity in the detector response.[PUBLICATION ABSTRACT]
ISSN:0543-1972
1573-8906
DOI:10.1023/A:1026177800820