Development of multiple-elution cartridge-based radioisotope concentrator device for increasing the super(99m)Tc and super(188)Re concentration and the effectiveness of super(99m)Tc/ super(99)Mo utilisation
A self-shielded, sterile and cartridge-based radioisotope concentrator device coupled to super(99m)Tc/ super(188)Re generators to increase the super(99m)Tc/ super(188)Re-concentration of the generator eluate was developed based on new aminoalkyl-functionalized silica sorbent which conditionally catc...
Gespeichert in:
Veröffentlicht in: | Journal of radioanalytical and nuclear chemistry 2015-02, Vol.303 (2), p.1173-1178 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A self-shielded, sterile and cartridge-based radioisotope concentrator device coupled to super(99m)Tc/ super(188)Re generators to increase the super(99m)Tc/ super(188)Re-concentration of the generator eluate was developed based on new aminoalkyl-functionalized silica sorbent which conditionally catches and release super(99m)Tc/ super(188)Re to concentrate the daughter nuclide of the generator eluates. The cartridge can be used for multiple elutions with an overall concentration factor of >100 and daughter nuclide recovery yield of >85 %. This device can be used for 10 days extension of super(99m)Tc-generator life-time, saving about 20 % of the generator activity and for "early" generator-elution programs, under which the generator is eluted at an optimized build-up time for increasing the effectiveness of super(99m)Tc/ super(99)Mo utilisation. |
---|---|
ISSN: | 0236-5731 1588-2780 |
DOI: | 10.1007/s10967-014-3439-9 |