Evaluation of acetabular cup initial fixation by using resonance frequency principle

The clinical practice shows that the loosening of acetabular cups is more frequent than stem loosening. With standard cups, the incidence of dislocation failure is highest in the first year after arthroplasty implantation. The aim of the study was to quantitatively evaluate the implant-bone stabilit...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine Journal of engineering in medicine, 2015-01, Vol.229 (1), p.3-8
Hauptverfasser: Henys, Petr, Capek, Lukas, Fencl, Jaroslav, Prochazka, Egon
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The clinical practice shows that the loosening of acetabular cups is more frequent than stem loosening. With standard cups, the incidence of dislocation failure is highest in the first year after arthroplasty implantation. The aim of the study was to quantitatively evaluate the implant-bone stability of a cementless acetabular cup prosthesis by using a device based on resonance frequency analysis. The evaluation of this device was done by finite element analysis and in vitro experiments. It was shown that not all the resonance frequencies can be measured by our device. The resonance frequencies vary within the range of 500–3000 Hz. The proposed power spectrum measurement gives the information about the absolute stiffness of the press-fit implant.
ISSN:0954-4119
2041-3033
DOI:10.1177/0954411914561485