Evaluation of MR issues for the latest standard brands of orthopedic metal implants: Plates and screws

Highlights • Although previous studies have indicated that most of the orthopedic implants are compatible in MR imaging system especially for titanium alloy, there are still concerns about the safety of patients with stainless steel implants, who were refused to a MR scan in most cases in our countr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of radiology 2015-03, Vol.84 (3), p.450-457
Hauptverfasser: Zou, Yue-fen, Chu, Bin, Wang, Chuan-bing, Hu, Zhi-yi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • Although previous studies have indicated that most of the orthopedic implants are compatible in MR imaging system especially for titanium alloy, there are still concerns about the safety of patients with stainless steel implants, who were refused to a MR scan in most cases in our country. • In this study, it was verified that both titanium alloy and stainless steel materials (plates and screws) cause a weak force and low MRI-related heating at a 1.5-T or less, which do not pose an additional hazard or risk to patients. In addition, we also had explored the influence of different sequences and parameters on size of metallic artifacts to obtain optimized pulse sequences with appropriate parameters for reducing artifacts, which would be convenient and useful in clinical work.
ISSN:0720-048X
1872-7727
DOI:10.1016/j.ejrad.2014.12.001