Explantation of the Novel Ahmed Glaucoma Valve M4 Implant

PURPOSE:To report a series of cases involving Ahmed Glaucoma Valve M4 (AGV) explantation and to discuss the surgical technique to remove the drainage device. METHODS:Four cases were identified that presented with AGV M4 postoperative complications necessitating tube shunt removal. Three patients pre...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of glaucoma 2015-02, Vol.24 (2), p.e1-e4
Hauptverfasser: Hu, Wanda D, Pro, Michael J, Fudemberg, Scott J, Moster, Marlene R
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:PURPOSE:To report a series of cases involving Ahmed Glaucoma Valve M4 (AGV) explantation and to discuss the surgical technique to remove the drainage device. METHODS:Four cases were identified that presented with AGV M4 postoperative complications necessitating tube shunt removal. Three patients presented with conjunctival erosion and 1 patient with persistent diplopia. AGV M4 implants were removed successfully between 1.5 and 9 months after implantation. RESULTS:Successful explantation of the AGV M4 novel implant was achieved in all cases without intraoperative or postoperative complications. CONCLUSIONS:If necessary, AGV M4 explantation can be successfully performed in the early postoperative period.
ISSN:1057-0829
1536-481X
DOI:10.1097/IJG.0000000000000192