What is the preferred strength setting of the sinogram-affirmed iterative reconstruction algorithm in abdominal CT imaging?
Our primary objective in this study was to determine the preferred strength setting for the sinogram-affirmed iterative reconstruction algorithm (SAFIRE) in abdominal computed tomography (CT) imaging. Sixteen consecutive clinical CT scans of the abdomen were reconstructed by use of traditional filte...
Gespeichert in:
Veröffentlicht in: | Radiological physics and technology 2015-01, Vol.8 (1), p.60-63 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Our primary objective in this study was to determine the preferred strength setting for the sinogram-affirmed iterative reconstruction algorithm (SAFIRE) in abdominal computed tomography (CT) imaging. Sixteen consecutive clinical CT scans of the abdomen were reconstructed by use of traditional filtered back projection (FBP) and 5 SAFIRE strengths:
S
1–
S
5. Six readers of differing experience were asked to rank the images on preference for overall diagnostic quality. The contrast-to-noise ratio was not significantly different between SAFIRE
S
1 and FBP, but increased with increasing SAFIRE strength. For pooled data,
S
2 and
S
3 were preferred equally but both were preferred over all other reconstructions.
S
5 was the least preferred, with FBP the next least preferred. This represents a marked disparity between the image quality based on quantitative parameters and qualitative preference. Care should be taken to factor in qualitative in addition to quantitative aspects of image quality when one is optimizing iterative reconstruction images. |
---|---|
ISSN: | 1865-0333 1865-0341 |
DOI: | 10.1007/s12194-014-0288-8 |