Design study of a Miniaturized Displacement Transducer (MDT) for an active middle ear implant system

People suffering from moderate to severe hearing loss can be treated with active middle ear implants. A new approach in this field is to implant an electromechanical transducer onto the round window membrane in order to improve coupling and be able to treat patients with middle-ear problems. In this...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomedical microdevices 2014-12, Vol.16 (6), p.805-814
Hauptverfasser: Kaltenbacher, D., Schächtele, J., Goll, E., Burkhardt, C., Arnold, H., Dalhoff, E., Zenner, H-P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:People suffering from moderate to severe hearing loss can be treated with active middle ear implants. A new approach in this field is to implant an electromechanical transducer onto the round window membrane in order to improve coupling and be able to treat patients with middle-ear problems. In this paper the design study for a miniaturized displacement transducer (MDT) for the round window is presented. Based on a requirement analysis, the basic principle and analytical modeling of the actuator is shown. A parameter variation study results in an optimized actuator configuration that is able to generate an amplification of 110 dB SPL theoretically. As a next step this actuator has to be manufactured and tested.
ISSN:1387-2176
1572-8781