Atrial defibrillation using low tilt pulses delivered by transcutaneous RF coupling

A new form of an atrial defibrillator is described, employing a transdermal transformer to couple a 7.2 MHz RF pulse to an implanted, passive receiver; this, in turn, delivers a low-tilt unipolar DC shock to the heart. In animal studies, cardioversion was 100% successful with pulses of 100 V amplitu...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Electronics letters 2001-11, Vol.37 (23), p.1-1
Hauptverfasser: Santos, J A, Manoharan, G, Evans, N E, Anderson, J McC, Kidawi, B J, Allen, J D, Adgey, A A J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new form of an atrial defibrillator is described, employing a transdermal transformer to couple a 7.2 MHz RF pulse to an implanted, passive receiver; this, in turn, delivers a low-tilt unipolar DC shock to the heart. In animal studies, cardioversion was 100% successful with pulses of 100 V amplitude and 10 ms width.
ISSN:0013-5194