Validation of a high performance liquid chromatographic method for alpha amanitin determination in urine
The objective of the present study was to develop and validate a liquid chromatographic method with electrochemical detection to measure alpha amanitin concentrations in urine after sample pretreatment with double mechanism (reversed phase/cation exchange) solid-phase extraction cartridges. The urin...
Gespeichert in:
Veröffentlicht in: | Forensic science international 1998-03, Vol.92 (1), p.59-68 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The objective of the present study was to develop and validate a liquid chromatographic method with electrochemical detection to measure alpha amanitin concentrations in urine after sample pretreatment with double mechanism (reversed phase/cation exchange) solid-phase extraction cartridges. The urine samples (10 ml) were purified and concentrated to 1 ml with elimination of matrix contaminants. The extracts were then separated by isocratic reversed-phase chromatography using a C18 column (4.6 mm×25 cm) with a mobile phase composed of 0.005 M phosphate buffer (pH 7.2) and acetonitrile (90:10). Coulometric detection was performed by applying an oxidation potential of +500 mV to a porous graphite electrode in a low-volume analytical cell. The limit of quantitation was 10 ng/ml with a signal-to-noise ratio=25. The linearity studied on spiked urine was satisfactory (
r=0.9966) from 10 ng/ml to 200 ng/ml. The average extraction recovery of alpha amanitin was 78%, determined using spiked urine samples ranging from 10–300 ng/ml. The intra-assay precision was checked at 10, 50 and 100 ng/ml levels (
n=10) in spiked urine samples, with resulting coefficients of variation of 3.6%, 2% and 1.5%, respectively. |
---|---|
ISSN: | 0379-0738 1872-6283 |
DOI: | 10.1016/S0379-0738(98)00006-1 |