Contamination of a purified water system by Aspergillus fumigatus in a new endoscopy reprocessing unit

Use of purified water for the final rinse stage of disinfected endoscopes is an important element of quality control. We describe the detection and management of Aspergillus fumigatus contamination of a new reverse osmosis unit supplying 10 automated endoscope reprocessor basins. Prompt detection an...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of infection control 2014-12, Vol.42 (12), p.1337-1339
Hauptverfasser: Khalsa, Kamaljit, BM, BSc, DRCOG, Smith, Andrew, BDS, FDSRCS, FRCPath, PhD, Morrison, Patrick, MIHEEM, BEng(Hons), Shaw, David, Peat, Marie, Howard, Paul, Hamilton, Kate, RGN, BN, MA(Hons), MPA, Stewart, Alan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Use of purified water for the final rinse stage of disinfected endoscopes is an important element of quality control. We describe the detection and management of Aspergillus fumigatus contamination of a new reverse osmosis unit supplying 10 automated endoscope reprocessor basins. Prompt detection and reaction to this contaminant were possible because of the introduction of a comprehensive program for microbiological monitoring of rinse waters, which included total viable counts, endotoxin, conductivity, and Pseudomonas spp.
ISSN:0196-6553
1527-3296
DOI:10.1016/j.ajic.2014.08.008