Correlation Between Knee Kinematics and Patellofemoral Contact Pressure in Total Knee Arthroplasty

Abstract The aim of this study is to evaluate the relationship between patellofemoral contact stress and intraoperative knee kinematic patterns after mobile bearing total knee arthroplasty (TKA). Medial osteoarthritic knees of forty-six posterior-stabilized total knee prostheses were evaluated using...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2014-12, Vol.29 (12), p.2305-2308
Hauptverfasser: Konno, Takuya, MD, Onodera, Tomohiro, MD, PhD, Nishio, Yusuke, MD, Kasahara, Yasuhiko, MD, PhD, Iwasaki, Norimasa, MD, PhD, Majima, Tokifumi, MD, PhD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract The aim of this study is to evaluate the relationship between patellofemoral contact stress and intraoperative knee kinematic patterns after mobile bearing total knee arthroplasty (TKA). Medial osteoarthritic knees of forty-six posterior-stabilized total knee prostheses were evaluated using a computed tomography-guided navigation system. Subjects were divided into two groups based on intraoperative knee kinematic patterns: the medial pivot group (n = 19) and the non-medial pivot group (n = 27). Mean intraoperative patello-femoral contact stress was significantly lower in the medial pivot group than in the non-medial pivot group (1.7 MPa vs. 3.2 MPa, P < 0.05). An intraoperative medial pivot pattern results in reduced patello-femoral contact stress.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2014.07.020