Single-shot calibration of soft x-ray mirrors using a sinusoidal transmission grating
Calibration of soft x-ray diagnostics is a challenge due to the lack of laboratory-size calibrated sources. An in situ calibration method for newly developed x-ray mirrors, is presented. The x-ray source is produced by laser-matter interaction, and twin transmission gratings which create two identic...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2014-11, Vol.85 (11), p.11E809-11E809 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Calibration of soft x-ray diagnostics is a challenge due to the lack of laboratory-size calibrated sources. An in situ calibration method for newly developed x-ray mirrors, is presented. The x-ray source is produced by laser-matter interaction, and twin transmission gratings which create two identical dispersion lines. The gratings have a sinusoidal transmission function, which produces a highly precise high-orders free spectrum. An x-ray mirror interacts with one of the dispersion lines, and the mirror efficiency curve as a function of wavelength is extracted. Mirror efficiency shows good agreement with the literature, and evidence of water layer may justify the need of in situ calibration. |
---|---|