Key comparison CCEM-K7: AC voltage ratio

We report the results of the international comparison of low-frequency ac voltage ratio: CCEM-K7. The participants made measurements of a unique travelling standard: an inductive voltage divider which provided the 20 ac voltage ratios chosen for the comparison. The nominal ratios chosen were: 0.1 to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Metrologia 2012-01, Vol.49, p.1-1
Hauptverfasser: Robinson, Ian, Belliss, Janet, Bryant, Stephen, Sánchez, Antonio, Álvarez, Yolanda, Schweiger, Kurt, Díaz, Carlos, Neira, Miguel, Callegaro, Luca, Lee, Rae Duk, Blanc, Isabelle, Overney, Frederic, He, XiaoBing, Ding, Cheng, Qian, ZhongTai, Waltrip, Bryan, Small, Greig, Fiander, John, Coogan, Peter, Johnson, Heather Leigh, Nakamura, Yasuhiro, Dierikx, Erik, Saxena, Anil Kishore, Saleem, Mohd, Wood, Barry, Ramm, Guenther, Eklund, Gunnar, Turhan, Enis, Semenov, Yuri
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report the results of the international comparison of low-frequency ac voltage ratio: CCEM-K7. The participants made measurements of a unique travelling standard: an inductive voltage divider which provided the 20 ac voltage ratios chosen for the comparison. The nominal ratios chosen were: 0.1 to 0.9, 0.01 and 1/11 to 10/11. Each of the 17 participants measured the in-phase and quadrature components of all 20 ratios at a frequency of 1 kHz, and 7 laboratories made additional, optional, measurements at a frequency of 55 Hz. The report consists of two separate parts: the first part describes the comparison and provides detailed uncertainty budgets for each participant; the second part describes the method used to analyse the results, gives the results of the comparison and tabulates the raw data provided by each participant.
ISSN:0026-1394
1681-7575