Development of a static headspace gas chromatographic procedure for the routine analysis of volatile fatty acids in wastewaters

An optimised procedure has been developed for the routine analysis of volatile fatty acids in wastewater matrices, using static headspace gas chromatography with flame ionisation detection. Factors such as sample volume, sample pre-treatment and the time and temperature of sample equilibration have...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Chromatography A 2002-02, Vol.945 (1), p.195-209
Hauptverfasser: Cruwys, J.A, Dinsdale, R.M, Hawkes, F.R, Hawkes, D.L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An optimised procedure has been developed for the routine analysis of volatile fatty acids in wastewater matrices, using static headspace gas chromatography with flame ionisation detection. Factors such as sample volume, sample pre-treatment and the time and temperature of sample equilibration have been included in an optimisation model designed to provide maximum detector response for acetic, propionic, iso- and n-butyric and iso- and n-valeric acids in the concentration range 0–1000 mg/l. Optimal headspace conditions were observed when equilibrating at 85 °C for 30 min, using a 2.0 ml sample volume with the addition of 1.0 ml of NaHSO 4 (62%, w/v) into standard 22.3 ml vials. 2-Ethylbutyric acid was used as an internal standard. The suitability of ordinary least squares regression and weighted least squares regression models for the purposes of calibration and quantification were investigated. A weighted least squares linear regression model applied to the heteroscedastic data provided lower detection limits, e.g. 3.7 and 3.3 mg/l for acetic and propionic acids.
ISSN:0021-9673
DOI:10.1016/S0021-9673(01)01514-X