Sonic field measurement using light computerized tomography

In this work, we propose a new method of sonic field measurement using light waves. The advantage of using light is to be able to measure sonic fields without contact. The main conventional method involving use of light is sound pressure intensity measurement using optical diffraction of sonic field...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:JPN J APPL PHYS PART 1 REGUL PAP SHORT NOTE REV PAP 1997, Vol.36 (5B), p.3199-3202
Hauptverfasser: YOSHIOKA, M, MIZUTANI, K, NAGAI, K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this work, we propose a new method of sonic field measurement using light waves. The advantage of using light is to be able to measure sonic fields without contact. The main conventional method involving use of light is sound pressure intensity measurement using optical diffraction of sonic fields. On the other hand, the new method can be used to measure amplitude and phase distributions of sound pressure as a kind of wave field by means of tomography. We perform a simulation assuming a circular ultrasonic transducer. Variations of light wave phase induced by generation of sonic fields are calculated in the simulation. Using the numerical results, we reconstruct the sonic field images. It is confirmed that images of an ultrasonic field can be obtained.
ISSN:0021-4922
1347-4065
DOI:10.1143/JJAP.36.3199