Simplified liquid chromatographic determination of glyphosate and metabolite residues in environmental water using post-column fluorogenic labelling

Glyphosate and its metabolite aminomethylphosphonic acid in environmental water can be preconcentrated with an anion-exchange column, eluted with potassium citrate solution and determined directly by liquid chromatography with a post-column reactor and a fluorescence detector. The limit of detection...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytica chimica acta 1990, Vol.233 (2), p.311-314
Hauptverfasser: Wigfield, Yuk Y., Lanouette, Monique
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Glyphosate and its metabolite aminomethylphosphonic acid in environmental water can be preconcentrated with an anion-exchange column, eluted with potassium citrate solution and determined directly by liquid chromatography with a post-column reactor and a fluorescence detector. The limit of detection and average recovery are 1 μg 1 −1 and 89.3% for glyphosate and 0.4 μg l −1 and 86.3% for the metabolite.
ISSN:0003-2670
1873-4324
DOI:10.1016/S0003-2670(00)83495-9