Added value of shear-wave elastography for evaluation of breast masses detected with screening US imaging
To evaluate the additional value of shear-wave elastography (SWE) to B-mode ultrasonography (US) and to determine an appropriate guideline for the combined assessment of screening US-detected breast masses. This study was conducted with institutional review board approval, and written informed conse...
Gespeichert in:
Veröffentlicht in: | Radiology 2014-10, Vol.273 (1), p.61-69 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate the additional value of shear-wave elastography (SWE) to B-mode ultrasonography (US) and to determine an appropriate guideline for the combined assessment of screening US-detected breast masses.
This study was conducted with institutional review board approval, and written informed consent was obtained. From March 2010 to February 2012, B-mode US and SWE were performed in 159 US-detected breast masses before biopsy. For each lesion, Breast Imaging Reporting and Data System (BI-RADS) category on B-mode US images and the maximum stiffness color and elasticity values on SWE images were assessed. A guideline for adding SWE data to B-mode US was developed with the retrospective cohort to improve diagnostic performance in sensitivity and specificity and was validated in a distinct prospective cohort of 207 women prior to biopsy.
Twenty-one of 159 masses in the development cohort and 12 of 207 breast masses in the validation cohort were malignant. In the development cohort, when BI-RADS category 4a masses showing a dark blue color or a maximum elasticity value of 30 kPa or less on SWE images were downgraded to category 3, specificity increased from 9.4% (13 of 138) to 59.4% (82 of 138) and 57.2% (79 of 138) (P < .001), respectively, without loss in sensitivity (100% [21 of 21]). In the validation cohort, specificity increased from 17.4% (34 of 195) to 62.1% (121 of 195) and 53.3% (104 of 195) (P < .001) respectively, without loss in sensitivity (91.7% [11 of 12]).
The addition of SWE to B-mode US improved diagnostic performance with increased specificity for screening US-detected breast masses. BI-RADS category 4a masses detected at US screening that showed a dark blue color or a maximum elasticity value of 30 kPa or less on SWE images can be safely followed up instead of performing biopsy. |
---|---|
ISSN: | 0033-8419 1527-1315 |
DOI: | 10.1148/radiol.14132443 |