Routine application of Raman spectroscopy in the quality control of hospital compounded ganciclovir

[Display omitted] This study compares the performance of a reference method of HPLC to Raman spectroscopy (RS) for the analytical quality control (AQC) of therapeutic objects. We assessed a model consisting of a widely used antiviral drug, i.e., ganciclovir, which was compounded in a medical device...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of pharmaceutics 2014-10, Vol.474 (1-2), p.193-201
Hauptverfasser: Amin, Alexandre, Bourget, Philippe, Vidal, Fabrice, Ader, Flavie
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:[Display omitted] This study compares the performance of a reference method of HPLC to Raman spectroscopy (RS) for the analytical quality control (AQC) of therapeutic objects. We assessed a model consisting of a widely used antiviral drug, i.e., ganciclovir, which was compounded in a medical device and then transferred in a vacuum glass vial prior to analyses. As the aim of the alternative RS method is to replace the destructive, time-consuming HPLC method, requiring sample preparation, it needs to be demonstrated that RS performs at least as good as the HPLC method. Therefore, the two methods were validated by calculating the accuracy profile and provided excellent results for the analytical validation key criteria, i.e., trueness, precision and accuracy, ranging from 0.8 to 10mg/mL. The Spearman and Kendall correlation tests (p-value
ISSN:0378-5173
1873-3476
DOI:10.1016/j.ijpharm.2014.08.028