Online solid phase extraction LC–MS/MS method for the analysis of succinate dehydrogenase inhibitor fungicides and its applicability to surface water samples
A sensitive and selective analytical method, based on online solid phase extraction coupled to LC–MS/MS, was developed and validated to determine traces of several recently introduced fungicides in surface water and wastewater. The list of target analytes included eight succinate dehydrogenase inhib...
Gespeichert in:
Veröffentlicht in: | Analytical and bioanalytical chemistry 2014-10, Vol.406 (25), p.6419-6427 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A sensitive and selective analytical method, based on online solid phase extraction coupled to LC–MS/MS, was developed and validated to determine traces of several recently introduced fungicides in surface water and wastewater. The list of target analytes included eight succinate dehydrogenase inhibitors (bixafen, boscalid, fluopyram, flutolanil, fluxapyroxad, isopyrazam, penflufen, and penthiopyrad), and two other fungicides with different modes of action, fenpyrazamine and fluopicolide. Detection and quantification limits in various matrices were in the range of 0.1 to 2 and 0.5 to 10 ng/L, respectively. Moderate signal suppression was observed in surface water (≤15 %) and wastewater (≤25 %) and was well compensated by the selected internal standard. The intra- and inter-day precisions were generally |
---|---|