Silica nanoparticle-based microfluidic immunosensor with laser-induced fluorescence detection for the quantification of immunoreactive trypsin

The purpose of this study was to develop a silica nanoparticle-based immunosensor with laser-induced fluorescence (LIF) as a detection system. The proposed device was applied to quantify the immunoreactive trypsin (IRT) in cystic fibrosis (CF) newborn screening. A new ultrasonic procedure was used t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical biochemistry 2014-10, Vol.463, p.31-37
Hauptverfasser: Seia, Marco A., Stege, Patricia W., Pereira, Sirley V., De Vito, Irma E., Raba, Julio, Messina, Germán A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to develop a silica nanoparticle-based immunosensor with laser-induced fluorescence (LIF) as a detection system. The proposed device was applied to quantify the immunoreactive trypsin (IRT) in cystic fibrosis (CF) newborn screening. A new ultrasonic procedure was used to extract the IRT from blood spot samples collected on filter papers. After extraction, the IRT reacted immunologically with anti-IRT monoclonal antibodies immobilized on a microfluidic glass chip modified with 3-aminopropyl functionalized silica nanoparticles (APSN–APTES-modified glass chips). The bounded IRT was quantified by horseradish peroxidase (HRP)-conjugated anti-IRT antibody (anti-IRT–Ab) using 10-acetyl-3,7-dihydroxyphenoxazine (ADHP) as enzymatic mediator. The HRP catalyzed the oxidation of nonfluorescent ADHP to highly fluorescent resorufin, which was measured by LIF detector, using excitation lambda at 561nm and emission at 585nm. The detection limits (LODs) calculated for LIF detection and for a commercial enzyme-linked immunosorbent assay (ELISA) test kit were 0.87 and 4.2ngml−1, respectively. The within- and between-assay variation coefficients for the LIF detection procedure were below 6.5%. The blood spot samples collected on filter papers were analyzed with the proposed method, and the results were compared with those of the reference ELISA method, demonstrating a potential usefulness for the clinical assessment of IRT during the early neonatal period.
ISSN:0003-2697
1096-0309