Precision and Reproducibility of Measurements of Human Corneal Thickness with Rapid Optical Low-Coherence Reflectometry (OLCR)

This study investigates the precision and intraindividual variability of a clinical optical pachometer based on low-coherence reflectometry, which was used to measure the central thickness of a human cornea . The instrument, attached to a slit lamp, is a single mode fiber optic based Michelson inter...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Biomedical Optics 1999-01, Vol.4 (1), p.152-156
Hauptverfasser: Bo¨hnke, Matthias, Masters, Barry R, Wa¨lti, Rudolf, Ballif, Julien J, Chavanne, Philippe, Gianotti, Ronald, Salathe´, Rene´ Paul
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study investigates the precision and intraindividual variability of a clinical optical pachometer based on low-coherence reflectometry, which was used to measure the central thickness of a human cornea . The instrument, attached to a slit lamp, is a single mode fiber optic based Michelson interferometer with a high repetition rate as previously described. The same operator performed ten sets of measurements on the same subject, each consisting of 20 consecutive scans, on each day for three consecutive days. By computing the means from every scan series, the thickness of the central cornea with optical pachometry was found to be (range 518-521) m on day 1, (range 519-521) m on day 2, and (range 523-525) m on day 3. The thickness values on day 3, where the subject suffered from a cold without clinical ocular involvement, were different from the two previous days one way analysis of variance). Optical low-coherence reflectometry measurements of corneal thickness can be performed with high precision of about 1 m and a high intra- and intersession reproducibility. ©
ISSN:1083-3668
1560-2281