Component Size Mismatch of Metal on Metal Hip Arthroplasty: An Avoidable Never Event

Abstract Size mismatch of components used in total hip arthroplasty is a serious, preventable patient safety incident of unknown prevalence as many cases are not detected. Component size mismatch was found in 11 cases (0.9%) at our retrieval centre. All cases of mismatch were not detected on plain r...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2014-08, Vol.29 (8), p.1629-1634
Hauptverfasser: Whittaker, Robert K., BSc, Hexter, Adam, BSc, MBBS, Hothi, Harry S., BEng, MSc, PhD, Panagiotidou, Anna, BSc, MBBS, MRCS, Bills, Paul J., BEng, MSc, PhD, Skinner, John A., MBBS, FRCS (Eng), FRCS (Orth), Hart, Alister J., MA, MD, FRCS (Orth)
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Size mismatch of components used in total hip arthroplasty is a serious, preventable patient safety incident of unknown prevalence as many cases are not detected. Component size mismatch was found in 11 cases (0.9%) at our retrieval centre. All cases of mismatch were not detected on plain radiograph during routine clinical follow up and blood metal ion levels were elevated above the MHRA action level of 7 ppb. Root cause analysis identified manufacturer, hospital and surgeon factors that need to be addressed to reduce the incidence of this avoidable clinical problem. Retrieval analysis is the only method of confirming size mismatch and is likely to be under-represented in National Joint Registries that record the indication for revision at the time of revision.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2014.03.008