Are ECG Welsh cup electrodes effectively cleaned?
Re-usable Welsh electrocardiograph (ECG)-electrodes are potential vehicles for cross infection. This study confirmed that in-use ECG-electrodes are frequently contaminated with organisms such as coagulasenegative staphylococci and Gram-negative bacilli. The efficacy of five cleaning procedures was e...
Gespeichert in:
Veröffentlicht in: | The Journal of hospital infection 1989-11, Vol.14 (4), p.325-331 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Re-usable Welsh electrocardiograph (ECG)-electrodes are potential vehicles for cross infection. This study confirmed that in-use ECG-electrodes are frequently contaminated with organisms such as coagulasenegative staphylococci and Gram-negative bacilli. The efficacy of five cleaning procedures was evaluated. Immersing the electrodes in water at 60 °C for one hour was the most effective method of decontamination tested, following challenge with a standardised suspension of
Staphylococcus saprophyticus. Significant contamination persisted following simple cleaning measures. It is suggested that this was promoted by the inadequate removal of electrode gel which provided a protective environment for microorganisms. |
---|---|