Expansion and malapposition characteristics after bioresorbable vascular scaffold implantation
Objectives This study sought to investigate the postdeployment expansion and malapposition characteristics of the bioresorbable vascular scaffold (BVS) in real‐world practice. Background The material construct of the BVS precludes overexpansion, with consequent potential for scaffold underexpansion...
Gespeichert in:
Veröffentlicht in: | Catheterization and cardiovascular interventions 2014-07, Vol.84 (1), p.37-45 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objectives
This study sought to investigate the postdeployment expansion and malapposition characteristics of the bioresorbable vascular scaffold (BVS) in real‐world practice.
Background
The material construct of the BVS precludes overexpansion, with consequent potential for scaffold underexpansion and malapposition. In metallic stents, these features are associated with an increased risk of adverse events, including stent thrombosis. The postdeployment characteristics of the BVS are yet to be described outside clinical trials, where implantation occurred in straightforward lesion subsets.
Methods
Data from 25 patients undergoing BVS implantation were analyzed. Optical coherence tomography (OCT) was performed both before and after intervention to assess plaque composition, scaffold expansion and strut apposition. Manufacturer's compliance charts were used to predict expected minimal scaffold diameter and area.
Results
OCT pullback (522.2 mm) was analyzed. Overall, BVS achieved 82.5 ± 8.7 and 79.8 ± 12.3% of predicted minimal stent diameter and cross‐sectional area (SCA), respectively, with expansion reduced in middle third of the scaffold (central SCA 76.7 ± 10.9% vs. noncentral SCA 81.5 ± 12.7%, P |
---|---|
ISSN: | 1522-1946 1522-726X |
DOI: | 10.1002/ccd.25378 |