MRI with cardiac pacing devices – Safety in clinical practice

Abstract Objectives The aim of this study was to introduce a single centre “real life” experience of performing MRI examinations in clinical practice on patients with cardiac pacemaker systems. Additionally, we aimed to evaluate the safety of using a dedicated safety protocol for these patients. Mat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European journal of radiology 2014-08, Vol.83 (8), p.1387-1395
Hauptverfasser: Kaasalainen, Touko, Pakarinen, Sami, Kivistö, Sari, Holmström, Miia, Hänninen, Helena, Peltonen, Juha, Lauerma, Kirsi, Sipilä, Outi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Objectives The aim of this study was to introduce a single centre “real life” experience of performing MRI examinations in clinical practice on patients with cardiac pacemaker systems. Additionally, we aimed to evaluate the safety of using a dedicated safety protocol for these patients. Materials and methods We used a 1.5 T MRI scanner to conduct 68 MRI scans of different body regions in patients with pacing systems. Of the cardiac devices, 32% were MR-conditional, whereas the remaining 68% were MR-unsafe. We recorded the functional parameters of the devices prior, immediately after, and approximately one month after the MRI scanning, and compared the device parameters to the baseline values. Results All MRI examinations were completed safely, and each device could be interrogated normally following the MRI. We observed no changes in the programmed parameters of the devices. For most of the participants, the distributions of the immediate and one-month changes in the device parameters were within 20% of the baseline values, although some changes approached clinically important thresholds. Furthermore, we observed no differences in the variable changes between MR-conditional and MR-unsafe pacing systems, or between scans of the thorax area and other scanned areas. Conclusion MRI in patients with MR-conditional pacing systems and selected MR-unsafe systems could be performed safely under strict conditions in this study.
ISSN:0720-048X
1872-7727