The development of an amperometric microbial biosensor using Acetobacter pasteurianus for lactic acid
A microbial biosensor, using Acetobacter pasteurianus cells and an oxygen electrode, was developed for the determination of lactic acid. The bacterial cells were retained on a nylon membrane and attached to the surface of the oxygen electrode. In view of response time, stability and sensitivity, the...
Gespeichert in:
Veröffentlicht in: | Journal of biotechnology 1989, Vol.10 (3), p.241-252 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A microbial biosensor, using
Acetobacter pasteurianus cells and an oxygen electrode, was developed for the determination of lactic acid. The bacterial cells were retained on a nylon membrane and attached to the surface of the oxygen electrode. In view of response time, stability and sensitivity, the biosensor performed best at 26°C and in pH 6 phthalate buffer containing magnesium sulfate. The activity of the retained cells was stable for approximately 170 h and was regenerable. The biosensor exhibited a hyperbolic response to both D- and L-lactic acid in the range of 10
−4 M to 25 × 10
−3 M. However, in the range 10
−4 M to 15 × 10
−4 M the response was linear. The microbial biosensor was applicable for detecting lactate concentration in yogurt and milk, since it was not sensitive to lactose, sucrose and glucose — three major components of such dairy products. |
---|---|
ISSN: | 0168-1656 1873-4863 |
DOI: | 10.1016/0168-1656(89)90068-0 |