Interlaboratory testing of Insent e-tongues

[Display omitted] The first interlaboratory testing of electronic taste sensing systems was performed within five participating centers, each working with the Insent (Insent Inc., Atsugi-Shi, Japan) e-tongue. Preparation of the samples for the comprised four experiments, shipping of the samples and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of pharmaceutics 2014-08, Vol.469 (2), p.228-237
Hauptverfasser: Pein, Miriam, Gondongwe, Xolani Dereck, Habara, Masaaki, Winzenburg, Gesine
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:[Display omitted] The first interlaboratory testing of electronic taste sensing systems was performed within five participating centers, each working with the Insent (Insent Inc., Atsugi-Shi, Japan) e-tongue. Preparation of the samples for the comprised four experiments, shipping of the samples and evaluation of the results was performed at the University of Duesseldorf. The sensitivity (in this case the difference between lowest and highest sensor response) and slope of the regression line values, obtained within Experiment 1 and 2, have been found to serve as applicable evaluation criterions for interlaboratory comparability. Modified sensor responses could be attributed to aged sensors, but did not influence the results of either Experiment 3, dealing with the evaluation of film formulations, or Experiment 4, dealing with the evaluation of minitablet formulations, in a great amount. Presented PCA Score and Loading Scatter Plots as well as Euclidean distance patterns based on the raw sensor responses confirmed the comparable performance of Insent e-tongues of the participating centers.
ISSN:0378-5173
1873-3476