A digital data acquisition framework for the Versatile Array of Neutron Detectors at Low Energy (VANDLE)
Neutron energy measurements can be achieved using time-of-flight (ToF) techniques. A digital data acquisition system was developed for reliable ToF measurements with subnanosecond timing resolution based on digitizers with 10ns and 4ns sampling periods using pulse shape analysis algorithms. A valida...
Gespeichert in:
Veröffentlicht in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2014-02, Vol.737, p.22-28 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Neutron energy measurements can be achieved using time-of-flight (ToF) techniques. A digital data acquisition system was developed for reliable ToF measurements with subnanosecond timing resolution based on digitizers with 10ns and 4ns sampling periods using pulse shape analysis algorithms. A validation procedure was developed to confirm the reliability. The response of the algorithm to photomultiplier signals was studied using a specially designed experimental system based on fast plastic scintillators. The presented developments enabled digital data acquisition systems to instrument the recently developed Versatile Array of Neutron Detectors at Low-Energy (VANDLE). |
---|---|
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/j.nima.2013.11.028 |