Disintegration rate, gamma-ray emission probabilities and metastable half-life measurements of super(67)Ga
The procedure for determining the super(67)Ga disintegration rate by a primary method is described. The proposed triple 4 pi beta - gamma coincidence system consists of a thin window gas-flow 4 pi proportional counter (PC) coupled to a NaI(Tl) scintillator and a HPGe crystal. Independent pulse heigh...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2014-05, Vol.87, p.126-131 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The procedure for determining the super(67)Ga disintegration rate by a primary method is described. The proposed triple 4 pi beta - gamma coincidence system consists of a thin window gas-flow 4 pi proportional counter (PC) coupled to a NaI(Tl) scintillator and a HPGe crystal. Independent pulse height and occurrence time information is provided for the three detector outputs by means of a Software Coincidence System. Separate spectrometry measurements with a n-type reversed electrode coaxial Ge detector (REGe) were performed for obtaining gamma-ray emission probabilities per decay. Accurate values of disintegration rate, gamma-ray emission probabilities and the metastable half-life were achieved. |
---|---|
ISSN: | 0969-8043 |