Critical remarks on gross alpha/beta activity analysis in drinking waters: Conclusions from a European interlaboratory comparison
The most common gross alpha/beta standard methods used for drinking water analysis are discussed, and sources of interferences are reviewed from a metrological point of view. Our study reveals serious drawbacks of gross methods on the basis of an interlaboratory comparison analyzing commercial miner...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2014-05, Vol.87, p.429-434 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The most common gross alpha/beta standard methods used for drinking water analysis are discussed, and sources of interferences are reviewed from a metrological point of view. Our study reveals serious drawbacks of gross methods on the basis of an interlaboratory comparison analyzing commercial mineral water samples with the participation of 71 laboratories. A proposal is made to obtain comparable measurement results using true standardized methods.
•Gross alpha/beta standard methods for drinking water analysis are discussed.•Large spread of results (up to 2 orders of magnitude) observed in comparisons.•Sources of interferences are reviewed.•We propose to use true standardized methods to obtain better measurement results. |
---|---|