Abstract Background In clinical settings, where arthrometers are mainly used by different clinicians, knowing the inter-rater reliability of the instrument is crucial in order for the results from different examiners to be accurately interpreted and limitations fully understood. The aim of this stud...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The knee 2014-03, Vol.21 (2), p.541-543
Hauptverfasser: Vauhnik, Renata, Morrissey, Matthew C, Perme, Maja Pohar, Sevsek, France, Rugelj, Darja
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Background In clinical settings, where arthrometers are mainly used by different clinicians, knowing the inter-rater reliability of the instrument is crucial in order for the results from different examiners to be accurately interpreted and limitations fully understood. The aim of this study was to evaluate the inter-rater reliability of the GNRB® knee arthrometer. Methods Knee anterior laxity in both knees was tested in a group of young, uninjured subjects (N = 27, 13 females) by two examiners. Knee anterior laxity was calculated at test forces of 134 N and 250 N with values presented for the unstandardised and standardised conditions (relative to patellar stabilisation force). Results The ICCs ranged from 0.220 to 0.424. Conclusions The inter-rater reliability of the GNRB® knee arthrometer is low.