Low-dose CT in body-packers: delineation of body packs and radiation dose in a porcine model
Purpose To compare low-dose computed tomography (CT) with standard CT and conventional radiography (CR) regarding delineation of body packs and radiation dose. Methods Nine samples of illicit drugs including cocaine, heroin, and hashish were positioned in the rectum of a 121.5 kg pig cadaver. Each s...
Gespeichert in:
Veröffentlicht in: | Forensic science, medicine, and pathology medicine, and pathology, 2014-06, Vol.10 (2), p.170-178 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To compare low-dose computed tomography (CT) with standard CT and conventional radiography (CR) regarding delineation of body packs and radiation dose.
Methods
Nine samples of illicit drugs including cocaine, heroin, and hashish were positioned in the rectum of a 121.5 kg pig cadaver. Each sample was scanned on a 64-row MDCT with 120 kV: one standard modulated pelvic protocol (STD), and without modulation at 80 mA (LD80), 30 mA (LD30), and 10 mA (LD10). Additionally, conventional abdominal anterior–posterior radiographs (77 kV and 106 ± 13 mA) were taken. Body pack characteristics (wrapping, content, shape) were rated independently by two radiologists and summarized to a delineation score from 0 to 9 with scores ≥6 representing sufficient delineation. Mean delineation scores were calculated for CR and CT protocols. These were additionally differentiated for readings in soft tissue (
S
), lung (
L
), user defined, variable window settings (
V
), and in cumulative window evaluation including all the other window settings (SLV). Effective doses were calculated (mSv).
Results
The CR delineation score was insufficient (3.1 ± 2.5; 2.4 ± 0.3 mSv). For CT, the SLV window setting performed best (
p
|
---|---|
ISSN: | 1547-769X 1556-2891 |