The calibration of a pixel sensor using both fluorescence and transmitted X-ray photons

Pixel sensors have been calibrated using both fluorescence X-ray photons and an X-ray beam obtained by the transmission technique. The X-rays were generated by an Amptek EDIX 40 X-ray tube (maximum voltage 40kV). During the fluorescence calibration the pixel sensor was placed in front of the target...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2013-08, Vol.718, p.336-338
Hauptverfasser: Bissi, L., Bizzaglia, S., Bizzarri, M., Farnesini, L., Menichelli, M., Meroli, S., Papi, A., Piluso, A., Saha, A., Scolieri, G., Servoli, L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Pixel sensors have been calibrated using both fluorescence X-ray photons and an X-ray beam obtained by the transmission technique. The X-rays were generated by an Amptek EDIX 40 X-ray tube (maximum voltage 40kV). During the fluorescence calibration the pixel sensor was placed in front of the target in an off-beam position; the resulting photons hitting the detector were emitted by fluorescence in all directions with an energy which is typical of the fluorescence lines of the target material. During the calibration in the transmission mode the detector was placed behind the target, acting now as a filter, and the energy of the photons was tuned by adjusting the voltage of the tube and the thickness of the target. In this paper the comparison between the two methods will be shown. From the results of this test, it is possible to infer that transmission is more efficient (higher photon yield) and flexible (more energy points are possible) but produces broader spectral lines while fluorescence has a better energy definition. A reasonable strategy to benefit from both methods is using fluorescence to calibrate a spectrometer that will be used to evaluate the energy of the X-rays emitted in the transmission mode. The results of this calibration will be shown in this paper.