An improved method for determining virucidal efficacy of a chemical disinfectant using an electrical impedance assay

A major problem with the testing of virucidal efficacy using current protocols is that scoring of virus-induced cytopathic effect (CPE) is dependent on subjective visual interpretation using light microscopy. The current report details the use of an electrical impedance assay (xCELLigence, ACEA Bios...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of virological methods 2014-04, Vol.199, p.25-28
Hauptverfasser: Ebersohn, Karen, Coetzee, Peter, Venter, E.H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A major problem with the testing of virucidal efficacy using current protocols is that scoring of virus-induced cytopathic effect (CPE) is dependent on subjective visual interpretation using light microscopy. The current report details the use of an electrical impedance assay (xCELLigence, ACEA Biosciences) for its utility in virucidal efficacy testing. In this study, the xCELLigence system was used in a procedure developed from guidelines given by the Deutsche Vereiniging zur Bekämpfung der Viruskrankheiten (DVV) (German Association for the Control of Virus Diseases) in order to demonstrate the inactivation of infectious bursal disease virus using a commercial virucide. Although the modified DVV assay using the xCELLigence system yielded identical results (i.e. a 5-log10 reduction in viral infectivity) as the traditional DVV assay, the system allows virucidal efficacy and cytotoxicity to be measured in a more precise and reproducible fashion.
ISSN:0166-0934
1879-0984
DOI:10.1016/j.jviromet.2013.12.023