Determination of pentachlorophenol in rabbit and human liver using liquid chromatography and gas chromatography with electron-capture detection

A method has been developed for the determination of ppb levels of pentachlorophenol (PCP) in liver by column liquid chromatography with on-line electron-capture detection (LC-ECD). The technique is compared with gas chromatography-ECD (GC-ECD) for the analysis of rabbit and human livers. Good resul...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Chemosphere (Oxford) 1988, Vol.17 (7), p.1301-1308
Hauptverfasser: Maris, F.A., de Jong, G.J., Somsen, G.W., Brinkman, U.A.Th, Goewie, C.E., van den Broek, H.H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method has been developed for the determination of ppb levels of pentachlorophenol (PCP) in liver by column liquid chromatography with on-line electron-capture detection (LC-ECD). The technique is compared with gas chromatography-ECD (GC-ECD) for the analysis of rabbit and human livers. Good results were generally obtained for LC-ECD down to the low-ppb range; however, in some cases problems were observed when the sample is not dissolved in the LC eluent. LC-ECD can adequately be used as a confirmation method for GC-ECD.
ISSN:0045-6535