Design and evaluation of two different finger concepts for body-powered prosthetic hand

The goal of this study was to find an efficient method of energy transmission for application in an anthropomorphic underactuated body-powered (BP) prosthetic hand. A pulley-cable finger and a hydraulic cylinder finger were designed and tested to compare the pulley-cable transmission principle with...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of rehabilitation research and development 2013-01, Vol.50 (9), p.1253-1266
Hauptverfasser: Smit, Gerwin, Plettenburg, Dick H, van der Helm, Frans C T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The goal of this study was to find an efficient method of energy transmission for application in an anthropomorphic underactuated body-powered (BP) prosthetic hand. A pulley-cable finger and a hydraulic cylinder finger were designed and tested to compare the pulley-cable transmission principle with the hydraulic cylinder transmission principle. Both fingers had identical dimensions and a low mass. The only thing that differed between the fingers was the transmission principle. The input energy was measured for a number of tasks. The pulley-cable finger required more input energy than the hydraulic cylinder finger to perform the tasks. This was especially the case in tasks that required high pinch forces. The hydraulic cylinder transmission is therefore the more efficient transmission for application in BP prosthetic fingers.
ISSN:0748-7711
1938-1352
DOI:10.1682/JRRD.2012.12.0223