Comparative Outcomes of Transvenous Extraction of Sprint Fidelis and Riata Defibrillator Leads: A Single Center Experience
Comparison of Extraction of Recalled Defibrillator Leads Introduction The FDA has issued class I advisories for Medtronic Sprint Fidelis® and St. Jude Medical RiataTM ICD lead families. Transvenous RiataTM ICD lead extraction is typically considered higher risk than Fidelis® extraction, based on lon...
Gespeichert in:
Veröffentlicht in: | Journal of cardiovascular electrophysiology 2014-01, Vol.25 (1), p.36-42 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Comparison of Extraction of Recalled Defibrillator Leads
Introduction
The FDA has issued class I advisories for Medtronic Sprint Fidelis® and St. Jude Medical RiataTM ICD lead families. Transvenous RiataTM ICD lead extraction is typically considered higher risk than Fidelis® extraction, based on longer duration from implant, presence of externalized conductors and lack of silicone backfill in the SVC and RV coils. However, published data comparing procedural outcomes between these leads are limited.
Methods
Records were reviewed for all patients undergoing transvenous extraction of Sprint Fidelis® or RiataTM ICD leads at the Vanderbilt Heart and Vascular Institute from July 2006 to April 2013 to ascertain indication for extraction, procedural details, complications, and 30‐day mortality.
Results
There were significant differences between those undergoing extraction of a Sprint Fidelis® (n = 145) or RiataTM lead (n = 47). In the RiataTM group, device‐related endocarditis was a more common indication for extraction, the mean duration of implant was longer, and larger excimer laser sheaths were required. Lead malfunction was a more common indication in the Fidelis® group. There were no statistically significant differences in median procedure duration, procedural success (97.9% vs 95.7%, P = 0.41), median length of hospital stay (1 day vs 1 day, P = 0.23), procedural complication rate (5.5% vs 10.6%, P = 0.23) or 30‐day mortality (2.1% vs 2.1%, P = 0.98). Analyses excluding patients with device infection revealed similar results.
Conclusion
Despite differences in baseline characteristics, this study indicates that Medtronic Sprint Fidelis® and St. Jude RiataTM ICD leads have similar procedural outcomes with transvenous lead extraction. |
---|---|
ISSN: | 1045-3873 1540-8167 |
DOI: | 10.1111/jce.12265 |