Humeral Head Reconstruction for Hill-Sachs Defects: A Biomechanical Comparison of 2 Fixation Techniques for Bone Grafting

Purpose The purpose of this biomechanical study was to compare anterograde with retrograde screw fixation for allograft reconstruction of Hill-Sachs defects. Methods In 8 pairs of fresh-frozen humeral heads, a 40% Hill-Sachs defect was created. The resultant wedge-shaped osteochondral fragment was u...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Arthroscopy 2014, Vol.30 (1), p.22-28
Hauptverfasser: Puskas, Gabor J., M.D, Giles, Joshua W., B.E.Sc, Degen, Ryan M., M.D, Johnson, James A., Ph.D, Athwal, George S., M.D., F.R.C.S.C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose The purpose of this biomechanical study was to compare anterograde with retrograde screw fixation for allograft reconstruction of Hill-Sachs defects. Methods In 8 pairs of fresh-frozen humeral heads, a 40% Hill-Sachs defect was created. The resultant wedge-shaped osteochondral fragment was used as allograft. For each technique, two 3.75-mm screws were used for fixation. To test the strength of fixation, a custom tool was used that would apply load to the graft. By use of a materials testing machine, a staircase cyclic loading protocol was performed (500 cycles at 10, 20, 30, and 40 N) and then load to failure. Graft displacement was measured by an optical tracking system. Results For the 2 techniques, graft displacement increased with increasing load and increasing number of cycles up to a mean of 0.9 ± 0.42 mm for anterograde fixation and 1.1 ± 0.79 mm for retrograde fixation. This increase was significant within each technique across all 4 loading levels ( P < .05). However, there were no significant differences in graft displacement between the 2 techniques at any loading level or number of cycles ( P  = .16 to P  = .96). In addition, the load to failure between the anterograde and retrograde techniques (98.5 N and 95.6 N, respectively) was not significantly different ( P  = .706). Conclusions The initial fixation and failure strength of anterograde and retrograde graft fixation techniques for substantial Hill-Sachs defects do not significantly differ in a biomechanical cadaveric model. Clinical Relevance This biomechanical study supports that in an engaging Hill-Sachs defect, both anterograde and retrograde screw fixation techniques can be used for fixation of humeral head allografts.
ISSN:0749-8063
1526-3231